BD recently initiated a voluntary recall (February 2017) for a subset of Research Use Only (RUO) single vial reagents with fluorophores. This recall was initiated after observing sporadic performance and vial-to-vial variability during routine internal testing as well as through complaint investigations. (Refer to recalled product formats in Table I)
BD has completed a thorough investigation and evaluation of product inventory, and based on those outcomes, the recall is being expanded to include additional lots of RUO single vial reagents with fluorophores. (Refer to additional product formats in Table II)
A comprehensive investigation was initiated to identify root cause and evaluate product inventory and distribution. As a result of this investigation, BD has confirmed that a prolonged exposure to high intensity broad spectrum LED lighting was the root cause that resulted in sporadic performance and vial-to-vial variation attributable to light induced degradation in certain lots of single vial reagents. This exposure occurred after a warehouse relocation in May 2016 where high intensity broad spectrum LED lighting had been installed to enhance employee working conditions. Product degradation from prolonged light exposure has the potential to result in reduced signal intensity that may cause changes in cell staining patterns, lower quantitative values for positive populations or no signal. In the case of tandem dyes, it may also result in changes to the values in the fluorescence spectral overlap matrix for the reagent.
Upon identification of the impact this lighting had on the RUO single vial reagents, BD promptly took steps to modify the lighting conditions to be equivalent to the previous warehouse. All testing to date has confirmed that our correction has been effective. Additional corrective actions are under evaluation to prevent recurrence of this situation in the future.
If you have purchased any of the recalled product lots on the Product Lookup we recommend you take the following actions:
BD is diligently working to replenish the supply chain with newly manufactured product. Therefore, many SKUs are available to ship upon receipt of your replacement order. Other SKUs will ship as soon as they are available.
| Product Format | Brand(s) |
|---|---|
| Alexa Fluor® 700 | BD Pharmingen™ BD Phosflow™ |
| APC-Cy™7 | BD Pharmingen™ |
| APC-H7 | BD Pharmingen™ |
| APC-R700 | BD Horizon™ |
| BUV661 | BD Horizon™ |
| BV605 | BD Horizon™ |
| Product Format | Brand(s) |
|---|---|
| BV650 | BD Horizon™ |
| BV711 | BD Horizon™ |
| BV786 | BD Horizon™ |
| PE-CF594 | BD Horizon™ |
| PE-Cy™5 | BD Pharmingen™ |
| Product Format | Brand(s) |
|---|---|
| Alexa Fluor® 488 | BD Pharmingen™ BD Phosflow™ |
| Alexa Fluor® 555 | BD Pharmingen™ |
| Alexa Fluor® 647 | BD Pharmingen™ BD Phosflow™ |
| APC | BD Pharmingen™ BD Phosflow™ |
| APC-H7 | BD Pharmingen™ BD Phosflow™ |
| APC-R700 | BD Horizon™ |
| BUV661 | BD Horizon™ |
| BV510 | BD Horizon™ BD Phosflow™ |
| BV650 | BD Horizon™ |
| Product Format | Brand(s) |
|---|---|
| BV711 | BD Horizon™ |
| BV750 | BD Horizon™ |
| BV786 | BD Horizon™ |
| Cy™5 | BD Pharmingen™ |
| Cy™5.5 | BD Pharmingen™ |
| PE | BD Pharmigen™ BD Phosflow™ BD™ Cytometric Bead Array (CBA) BD™ DimerX |
| PE-CF594 | BD Horizon™ BD Phosflow™ |
| PE-Cy™7 | BD Pharmingen™ BD Phosflow™ |
| PerCP-Cy™5.5 | BD Pharmingen™ BD Phosflow™ |
Note: Certain lots from each format were recalled. For your specific information from Table II please refer to your account specific recall letter or search using the Product Lookup.