Letters with a customer-specific list of purchased products that are part of the product notification are being mailed to customers. If you believe you have received product subject to the product notification, but have not yet received a letter, you can determine whether your product is affected by using the Product Lookup or reaching out to your local sales representative for more information.
No. This product notification is specific to a subset of Research Use Only (RUO) single vial reagents with fluorophores identifiable through the Product Lookup.
If you have purchased any of the product lots affected by the product notification on the Product Lookup, we recommend that you take the following actions:
BD is diligently working to replenish the supply chain with newly manufactured product. Therefore, many SKUs are available to ship upon receipt of your replacement order. Other SKUs will ship as soon as they are available.
If you wish to continue to use your reagents that are part of the product notification, we recommend conducting QC testing using biological cells per your standard protocol on each vial to confirm that the reagent meets your performance expectations and reviewing any data generated by your research that utilized those reagents.
Based on our investigation, using beads to QC these vials can result in a false-positive result not seen with biological cells.
Due to the sporadic performance and vial-to-vial variability of the reagents part of the product notification, we recommend that you review your data because the product degradation has the potential to result in reduced signal that may cause changes in cell staining patterns, lower quantitative values for positive populations or no signal. In the case of tandem dyes, it may also result in changes to the values in the fluorescence spectral overlap matrix for the reagent. If you have additional questions, please contact your local BD Technical Application Specialists for more information.
Based on the investigation, the current storage requirement on the product labeling, “Store undiluted at 4°C and protected from prolonged exposure to light,” is sufficient to protect the product from degradation. If you have additional questions, please contact your local BD Technical Application Specialists for more information.
Other RUO single vial reagents with fluorophores were stored at the same warehouse while the high intensity broad spectrum LED lighting was in use. These product formats are not part of the product notification because we conducted extensive testing of these product formats and determined with a high level of statistical confidence and reliability that they will perform as expected. For testing, single vial reagent samples were obtained from throughout the highest exposure risk areas in our storage facility. Functional flow cytometry was used to evaluate mean fluorescence intensity (MFI), as well as changes to the values in the fluorescence spectral overlap matrix for tandem formats, comparing samples with their respective Quality Assurance retains. Additionally, accelerated stability studies on a sampling of these product formats confirmed that the labeled expiration dates remain valid. These product formats have been released and will continue to be distributed.
If you have additional questions, please contact your local BD Technical Application Specialists for more information.