United States

Recall Letters

April 2017

February 2017

FAQs

How do I know if I have recalled products?

Notifications with a customer-specific list of purchased products that are being recalled are being mailed to customers. If you believe you have received product subject to the recall, but have not yet received a letter, you can determine whether your product is being recalled by using the Product Lookup or reaching out to your local sales representative for more information.

Does this recall affect any other BD reagents?

No. This recall is specific to a subset of Research Use Only (RUO) single vial reagents with fluorophores identifiable through the Product Lookup.

What should I do if I have recalled product?

If you have purchased any of the recalled product lots on the Product Lookup, we recommend that you take the following actions:

  1. Discard any remaining recalled product inventory and request replacement, or
  2. Until replacement products are available, conduct QC testing using biological cells per your standard protocol on each vial to confirm that the reagent meets your performance expectations, and
  3. Review any data generated by your research that utilized recalled products.

BD is diligently working to replenish the supply chain with newly manufactured product. Therefore, many SKUs are available to ship upon receipt of your replacement order. Other SKUs will ship as soon as they are available.

I haven't noticed a problem with the recalled products I have been using; can I continue to use them?

If you wish to continue to use your recalled products, we recommend conducting QC testing using biological cells per your standard protocol on each vial to confirm that the reagent meets your performance expectations and reviewing any data generated by your research that utilized those reagents.

Why can't I use beads to QC my affected vials? Why do I have to QC on biological cells?

Based on our investigation, using beads to QC these vials can result in a false-positive result not seen with biological cells.

My recalled products have already been consumed in my research; how do I know if my work was adversely affected?

Due to the sporadic performance and vial-to-vial variability of the recalled products, we recommend that you review your data because the product degradation has the potential to result in reduced signal that may cause changes in cell staining patterns, lower quantitative values for positive populations or no signal. In the case of tandem dyes, it may also result in changes to the values in the fluorescence spectral overlap matrix for the reagent. If you have additional questions, please contact [email protected], 1-866-979-9408 or your local BD Technical Application Specialists for more information.

How do I know if the lighting conditions in my work environment may adversely affect the product?

Based on the investigation, the current storage requirement on the product labeling, "Store undiluted at 4˚C and protected from prolonged exposure to light," is sufficient to protect the product from degradation. If you have additional questions, please contact [email protected], 1-866-979-9408 or your local BD Technical Application Specialists for more information.

Additional information on tested and released products

Other RUO single vial reagents with fluorophores were stored at the same warehouse while the high intensity broad spectrum LED lighting was in use. These product formats are not being recalled because we conducted extensive testing of these product formats and determined with a high level of statistical confidence and reliability that they will perform as expected. For testing, single vial reagent samples were obtained from throughout the highest exposure risk areas in our storage facility. Functional flow cytometry was used to evaluate mean fluorescence intensity (MFI), as well as changes to the values in the fluorescence spectral overlap matrix for tandem formats, comparing samples with their respective Quality Assurance retains. Additionally, accelerated stability studies on a sampling of these product formats confirmed that the labeled expiration dates remain valid. These product formats have been released and will continue to be distributed.

If you have additional questions, please contact [email protected], 1-866-979-9408 or your local BD Technical Application Specialists for more information.